Dosimetric verification of Monte Carlo and Collapsed Cone Convolution (CCC) algorithms
1 Aleksandër Xhuvani University Elbasan, Albania,
2 Aleksandër Xhuvani University Elbasan, Albania,
3 University of Tirana, Albania,
4 Polytechnic University of Tirana, Albania,
5 Aleksandër Xhuvani University Elbasan, Albania
* This email address is being protected from spambots. You need JavaScript enabled to view it.
Published online: 15 April 2026
Abstract
Monte Carlo and Collapsed Cone Convolution (CCC) algorithms are widely used for dosimetric verification and calibration of linear accelerators in external beam radiotherapy. This study presents a comparative dosimetric verification of both algorithms at different depths and field sizes using an Elekta linear accelerator. The goal is to assess the agreement between theoretical dose distributions and measured data, a crucial aspect in ensuring treatment accuracy in clinical radiotherapy. Measurements were conducted using a photon beam, a Farmer-type ionization chamber, and a water phantom. The experimental results were compared to dose calculations obtained from the Monte Carlo (MC) and Collapsed Cone Convolution (CCC) algorithms. These algorithms provide a predictive framework for estimating radiation dose distribution in tissues at varying depths. Verifying their accuracy is essential for confirming the consistency between theoretical planning and physical dose delivery. Our results show a generally acceptable level of agreement between the measured and calculated doses, supporting the validity of both algorithms for clinical use. Differences observed in specific field sizes and depths offer insight into algorithm limitations and potential areas for optimization. This verification contributes to improved confidence in treatment planning systems and enhances the quality assurance process in modern radiotherapy.
Key words: Dosimetric verification / Monte Carlo algorithm / Collapsed Cone Convolution algorithm / Radiotherapy / Linear accelerator
© The Authors, published by EDP Sciences, 2026
This is an Open Access article distributed under the terms of the Creative Commons Attribution License 4.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.