Coincidence resolution time of two small scintillators coupled to high quantum-efficiency photomultipliers in a PET-like system
1 Dipartimento Interateneo di Fisica “M. Merlin”, Bari, Italy
2 Istituto Nazionale di Fisica Nucleare, Sezione di Bari, Bari, Italy
3 Istituto Nazionale di Fisica Nucleare, Sezione di Roma, Roma, Italy
4 National Centre for Nuclear Research (NCBJ) - Otwock - Świerk, Poland
5 Istituto Nazionale di Fisica Nucleare, Sezione di Lecce, Italy
a e-mail: roberto.perrino@le.infn.it
Published online: 20 March 2014
The lower limit of the time resolution for a positron emission tomography (PET) system has been measured for two scintillator types, LYSO:Ce and LuAG:Pr. Small dimension crystals and ultra bi-alkali phototubes have been used in order to increase the detected scintillation photons. Good timing resolutions of 118 ps and 223 ps FWHM have been obtained for two LYSO and two LuAG, respectively, exposed to a 22Na source.
© Owned by the authors, published by EDP Sciences, 2014
This is an Open Access article distributed under the terms of the Creative Commons Attribution License 2.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.