Issue |
EPJ Web of Conferences
Volume 66, 2014
INPC 2013 – International Nuclear Physics Conference
|
|
---|---|---|
Article Number | 10016 | |
Number of page(s) | 4 | |
Section | Nuclear Physics Based Applications | |
DOI | https://doi.org/10.1051/epjconf/20146610016 | |
Published online | 20 March 2014 |
https://doi.org/10.1051/epjconf/20146610016
Adaptation and security validation of an irradiator suitable for use of cesium-137 sources
1 Department of Physics, Federal University of Sergipe – UFS, São Cristóvão, Sergipe, Brazil
2 Federal Institute of Bahia – IFBA, Vitória da Conquista, Bahia, Brazil
a e-mail: rmv.fisica@gmail.com
Published online: 20 March 2014
With the aim of offering a new application for 137Cs sources that were previously used in brachytherapy, we developed an irradiator for TL and OSL dosimeters by adapting a shielded storage container; this irradiator was named SOPHIA. It was assembled from a cubic iron casing, filled with lead, with external dimensions of 25 × 25 × 25 cm3. It has three different drawers with two sample holders suitable for samples or dosimeters with dimensions of up to 2 cm. Dose distribution of the surface of sample holder was obtained with EBT-3 radiochromic film and detectors thermoluminescent CaSO4: Dy. Regarding the security conditions of irradiation, the whole routine procedure for placement and removal of samples was simulated using Monte Carlo code. The equipment presented proper security conditions and the doses received by users during procedures for placement and extraction of the samples are in accordance with international standards.
© Owned by the authors, published by EDP Sciences, 2014
This is an Open Access article distributed under the terms of the Creative Commons Attribution License 2.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.