Issue |
EPJ Web of Conferences
Volume 106, 2016
ISRD 15 – International Symposium on Reactor Dosimetry
|
|
---|---|---|
Article Number | 06005 | |
Number of page(s) | 10 | |
Section | Benchmarks and Intercomparisons | |
DOI | https://doi.org/10.1051/epjconf/201610606005 | |
Published online | 03 February 2016 |
https://doi.org/10.1051/epjconf/201610606005
The EWGRD Round Robin Measurement Exercise
1 AMEC, Clean Energy – Europe, Gloucester, UK
2 SCK•CEN, Mol, Belgium
3 C2TN and LEN, Instituto Superior Técnico, Universidade de Lisboa, Portugal
4 CEA, Cadarache, France
5 Budapest University of Technology and Economics, Hungary
6 RCŘ, Řež, Czech Republic
7 NRG, Petten, The Netherlands
8 CEA, Valduc, France
9 VTT, Finland
a Corresponding author: dean.thornton@amec.com
Published online: 3 February 2016
This paper presents the results of a round robin exercise carried out to compare specific activity measurements performed by eight European organisations on a set of ten neutron activation detectors containing the radio-nuclides 110mAg, 60Co, 54Mn, 46Sc and 94Nb. The purpose of the exercise was to demonstrate the level of consistency between the participating organisations in blind tests of measurements relevant to reactor metrology. The samples used were selected from a stock of pre-existing irradiated material held at SCK•CEN. Taking turns over a period of approximately 9 months, the participating organisations received the samples, measured them and provided their results to an independent referee who collated and compared the data. The inter-comparison has demonstrated good agreement between the participants with standard deviations for each dosimeter varying between 1.6% and 3.1%. The paper provides results of the EWGRD Round Robin in an anonymised form together with discussion and conclusions which may be drawn from the exercise.
© Owned by the authors, published by EDP Sciences, 2016
This is an Open Access article distributed under the terms of the Creative Commons Attribution License 2.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.