Calibration of PVDF Film Transducers for the Cavitation Impact Measurement
Technical University of Liberec, Dep. of Power Engineering Equipment, Studentska 2, 461 17 Liberec, Czech Republic
* Corresponding author: jan.hujer@tul.cz
Published online: 4 June 2018
This paper describes investigation of the influence of the protective layer thickness on the calibration sensitivity of PVDF films sensors for the cavitation impacts measurements. The PVDF film sensor is casted into an aluminium block. The drop ball method is used for the measurement of the relation between impact force and the voltage detected on the PVDF film sensor. The calibration constants are measured for three different protective layers thicknesses. Five different ball weights for 400 mm drop height are used to reach the required impact force range. The ball positions for the evaluation of the impact force are measured with a high speed camera. The voltage signal detected on the PVDF film clamps was measured with a high speed digitizer. The measured signals are analysed in LabVIEW Signal Express.
© The Authors, published by EDP Sciences, 2018
This is an Open Access article distributed under the terms of the Creative Commons Attribution License 4.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited. (http://creativecommons.org/licenses/by/4.0/).
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.