Scanning of a Double-Sided Germanium Strip Detector
1
Department of Physics, Indian Institute of Technology Ropar, Rupnagar – 140 001, Punjab, India
2
Department of Nuclear and Atomic Physics, Tata Institute of Fundamental Research, Homi Bhabha Road, Mumbai – 400 005, India
3
GSI Helmholtzzentrum für Schwerionenforschung GmbH, Planckstrasse 1, 64291 Darmstadt, Germany
4
University of Massachusetts Lowell, Lowell, Massachusetts 01854, USA
5
Department of Physics, North Campus, University of Delhi, Delhi – 110 007, India
palit@tifr.res.in, 2017phz0005@iitrpr.ac.in
Published online: 19 November 2021
This work presents the results from the characterization of a Position-Sensitive Planar Germanium (PSPGe) detector. The PSPGe detector is a double-sided orthogonal strip detector consisting of 10x10 electrical segmentation along the horizontal and vertical directions. The characterization was performed using the coincidence setup between the PSPGe detector and the well-characterized scanning system employing the positron annihilation correlation principle. The scanning system consists of a Position Sensitive Detector (PSD) and 22Na positron source. The main objective of this study is to deploy PSPGe detector for future decay experiments at the Facility for Antiproton and Ion Research (FAIR), Germany. The measurements have been performed to find the depth of gamma-ray interaction in the planar segmented detector. The 2-Dimensional image obtained from the PSD has been used to find the depth of gamma-ray interaction in the planar strip detector using pulse shape analysis. In addition, the sensitivity of PSPGe detector has been investigated by calculating the rise-time from pulse shapes for the front and back strips of the detector.
Key words: Double-Sided Strip Detector / Positron Annihilation Correlation / Position Sensitive Detector / Pulse Shape Analysis
© The Authors, published by EDP Sciences, 2021
This is an Open Access article distributed under the terms of the Creative Commons Attribution License 4.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.
Initial download of the metrics may take a while.