Issue |
EPJ Web of Conferences
Volume 77, 2014
16th International Congress of Metrology
|
|
---|---|---|
Article Number | 00014 | |
Number of page(s) | 6 | |
DOI | https://doi.org/10.1051/epjconf/20147700014 | |
Published online | 19 August 2014 |
https://doi.org/10.1051/epjconf/20147700014
Inductive voltage divider calibration with sampling method
Federal Institute of Metrology METAS, Lindenweg 50, 3003 Bern-Wabern, Switzerland
a Corresponding author: david.corminboeuf@metas.ch
Published online: 19 August 2014
This paper presents an automatic system developed at METAS for the calibration of inductive voltage dividers (IVD). The used method is primary for the n/10 ratio. With the described improvements, the calibration of IVDs has been made easier, faster and more reproducible than with the manual system used before. The uncertainties which can be reached with this new system are 28·10-9 for the in phase and quadrature parts at a frequency of 1 kHz.
© Owned by the authors, published by EDP Sciences, 2014
This is an Open Access article distributed under the terms of the Creative Commons Attribution License 2.0, which permits unrestricted use, distribution, and reproduction in any medium, provided the original work is properly cited.
Current usage metrics show cumulative count of Article Views (full-text article views including HTML views, PDF and ePub downloads, according to the available data) and Abstracts Views on Vision4Press platform.
Data correspond to usage on the plateform after 2015. The current usage metrics is available 48-96 hours after online publication and is updated daily on week days.